This state-of-the-art aerosol monitoring system from GRIMM is
the latest instrument in our renowned line of high speed counters. With a resolution of 1
count/liter, the aerosols are collected on a removable 47 mm Teflon filter for gravimetric
or chemical analysis. This unique dual technology is incorporated in all GRIMM
real-time particle size analyzers. The results can also be analyzed by our
1.177 GRIMM Software
program and displayed as:
-
particle counts in 15 channels
-
mass distribution in 15 classes
-
environmental PM-10, PM-2.5 and PM-1
-
occupational as inhalable, thoracic, respirables (alveolics)
-
temperature, humidity and wind speed changes
A
random sampling head collects the dust in accordance to the Johannesburger
Convention and leads the particles directly into the optical chamber with
the laser. There each particle is counted and classified by size and the
concentration is proportional to the collected time. Results can be
continuously given in one second or longer, up to one minute in intervals.
For use outside, a special weather housing was developed so the unit
can be used in rain, but without being sensitive to the humidity (particle
condensation). This system does not heat the sample pipe as heating the
sample pipe will simply drive out all the semi volatiles from the air and
lead to wrong results.
Key Features
Each
GRIMM Model 1.108 is hand
crafted one at a time to the highest engineering standards and in accordance
to the most current ISO 9000 requirements. Like all our 1.100 series
instruments, the Model 1.108 has optical technology for real-time monitoring
as well as a precise volume controlled air flow with an integrated PTFE
filter, so all the dust measured optically can be analyzed at a later date
for the composition and the gravimetric mass correlation.
Some
of the additional features:
-
Displays real time data in as little as six seconds.
-
Portable, weighing only 2.4 kg (5.4 lbs.) with re-chargeable battery.
-
15 different size channels
from 0.30 µm
to 20 µm
-
LCD displays results in particle counts or mass distribution.
-
Auto zeros and performs system self-diagnostics at the start of each
analysis.
-
Removable 47 mm PTFE filter for optional gravimetric and/or chemical
analysis of residue.
-
Removable data storage cards. Data cards can hold up to one year of data.
-
Warning limits can be preprogrammed into the unit with an audible alarm.
-
RS 232 serial port for optional computer interface.
-
Analog inputs for optional environmental sensors.
|
|
Product Specifications
 |
|
Particle Size |
0.30/0.40/0.50/0.65/0.80/1.0/1.6/2.0/3.0/4.0/
5.0/7.5/10/15/20 µm |
|
Count Range |
1 to 2,000,000 counts/liter |
|
Mass Range |
.1 to 100,000 µg/m3 |
|
Size channels |
16 channel sizes, mass in µg/m3 and particle counts/liter |
|
Sensitivity |
1 particle/liter |
|
Sample Flow Rate |
1.2 liters/minute, flow controlled |
|
Reproducibility |
+/- 2% |
|
PTFE Filter Size |
47mm |
|
Operating Temperature Range |
4oC to 45oC |
|
Physical Parameters
|
|
Dimensions |
24 x 12 x 6cm
(9.5 x 5 x 2.5") |
|
Weight |
2.4 kg (5.4 lbs.) |
|
Input/Output |
lx RS 232, 3x Analog |
|
Power Requirements |
Battery or 110/220 VAC with external power supply |
|
The following options and accessories are available:
-
Omni-directional, lsokinetic and clean room Sampling Heads
-
Stainless Steel or Poly-Resin Housing for outdoor operation
-
Temperature, Humidity and Windspeed sensors.
Software Options and Data Presentation
The standard LCD readings are:
-
Counts in particles/liter
-
Mass as µg/m3
-
Date and Time
-
Up to three sensor inputs
-
Location
-
Battery Status
-
Filter Status
-
Various system diagnostic and instrument status.
GRIMMs software is fully compatible with WINDOWS 95
, 2000, ME, XP and NT. Data can be presented in the following formats:
Environmental Mode: PM 10, PM 2.5 and PM
1 conventions.
Occupational Mode: lnhalable, Thoracic and Respirable (alveolic).
Mass mode: µg/m3 in
15 size channels
Particle counts: particles/liter in 15 sizes
Our software also offers options for complete statistical analysis of data as well as instrument performance and
complete system diagnostics. |
|