The primary application for such measurements is in the
exhaust of internal combustion engines.
To measure aerosol size distributions, this
instrument system performs three operational functions:
charging, size classification, and detection. Particle
charging is accomplished by exposing the aerosol sample
to intense monochromatic UV light which results in
photoelectric ejection of electrons from particle
surfaces. The charged aerosol then passes into the
classification section of the TR-DMPS, where particles
are separated into different size ranges via electrical
mobility classification. Finally, particles are
collected and detected by measurement of the electrical
currents associated with the net charge on the particle.
The Model
5.600
uses a bipolar charger to charge the particles. The
charged particles enter the 10 DMA columns where they
are classified by size.
From 10 DMA voltages 21 size channels can be obtained
continuously. These voltages can be altered depending on
the user specifications. The number of particles is
detected by highly sensitive faraday cup electrometers (FCE).
The Engine Exhaust Spectrometer (TR-DMPS), Model 5.600,
measures sub micrometer particles in the range from 7 to
630nm in electrical mobility diameter. The unique design
of this instrument enables it to provide fast particle
size distribution measurements. It is intended for
measuring engine exhaust particles and even low
environmental concentrations. This
system is a combination of up to 10 Vienna Type DMA-FCE (Faraday Cup Electrometer) steps. This allows measurements of particle distributions of 3 - 600 nm with a measuring time resolution of 5 Hz. The combination of up to 10 independent and
very sensitive fractioning and measuring units allows an
accuracy which had not been achieved until now.
It is specially
designed for measurements in the field of motor testing
and fast atmospheric changes. Its resolution and
consistency make this this system especially effective
at measuring transient driving cycles for automotive
emission measurements. Environmental monitoring of
fast changing aerosols like in nucleation events or
on-road and drive by tests can also be tested. The
instrument can operate under different pressure
conditions so it is suitable for measurements in front
of filters and traps.
In the standard operation mode, the TR-DMPS
reports on 21 channels. The DMA cluster allows the user
to measure with 21 channels from 5 to 630 nm (3
selectable size ranges) with a time resolution of 5 Hz.
Every 200 ms the instrument displays a new size
distribution over all channels. The internal data
acquisition reaches up to 100 Hz.
A built in sampling and dilution unit gets a
representative and stable sample aerosol to the sensor
for high concentration measurements from the raw engine
exhaust. A bipolar diffusion charger places a
predictable charge on the particles. Additionally, a
cyclone restricts inlet particles to the size range that
can be measured.
 |
FCE (Faraday Cup Electrometer)
The fast
resolution can be achieved by the highly sensitive and
fast FCE units. Each FCE is separate and has a
real response time on aerosol concentration changes
within 20 ms and a concentration sensitivity of 0.5 fA
for this response time. Our design of the FCEs
avoids the use of PTFE (PolyTetraFluoroEthylene)
and has an integrated isolator which is far less
sensitive to mechanical, pressure, or temperature stress
than other existing solutions.
Maintenance
Due to the physical separation of the different size
channels and the separated DMAlFCE units, the system
is very robust. For maintenance each DMA can be
cleaned individually. The electrometers are designed
for long term operation without any maintenance.
Reference System
Standard CPC and SMPS measurement systems are normally
calibrated or referenced against DMAI FCE-combinations.
A DMAlFCE is the established reference method for
validation of any other aerosol measurement system in
the nanometer range. Our DMA/FCE system is the basic
measurement method assuring highest resolution and
reliability. The GRIMM TR-DMPS, built up with 10 DMAlFCE
combinations is an absolute measurement system assuring
highest size resolution combined with quickest
resolution for fast measurements.
A calibration of this system is not needed. Any other
measurement principle will be calibrated and validated
with this unit.
Results and Software
Data is collected from
all FCEs at a rate of 100 Hz (100 times per second, the
total response time of the extremely sensitive unit is
200 ms. This leads with a data reduction algorithm to
complete particle size distribution measurements with 5
Hz. Real-time data display is shown 5x per second and
can be averaged during data collection.
TR-DMPS software is user
friendly and full-featured. It includes a complete data
inversion algorithm and allows data to be displayed and
exported in many different ways to meet a variety of
needs. It also has a three dimensional representation of
size distribution over time. The software collects data
with 5x per second and can display averaged data with
0.5 to 60 seconds averages.
Pull-down menus and
dialog boxes simplify setup and operation. Tests may be
started by a few simple mouse clicks, at a specific
time, or by an external trigger.
|
Fast Engine
Exhaust Spectrometer |
|
Particle size range
in nm (pre-selectable) |
10-630/ 7-365/ 5-225 |
|
Particle size
resolution |
21 channels |
|
DMA systems |
10 units |
|
Electrometer (FCE) |
10 units |
|
Charger mode of
operation |
bipolar (diffusion) |
|
Inlet cyclone 50%
cut point |
650 nm |
|
Time resolution |
5 Hz |
|
Flow rates: |
|
|
Aerosol flow rate
(per DMA) |
1.6 L/min |
|
Sheath airflow (per
DMA) |
8 L/min |
|
Dilution ratio |
1 : 7 to 1 :40 |
|
Inlet aerosol
temperature |
10 to 70°C |
|
Operating
temperature |
5 to 50 °C |
|
Storage temperature |
20 to 50°C |
|
Pressure range |
500 to 2000 mbar |
|
Humidity
(non-condensing) |
0-95% RH |
|
User interface |
Computer |
|
Instrument status
report |
LED's |
|
Data averaging |
0.5 to 60 sec |
|
Communications |
9-pin RS-232, TCP IP |
|
Electrical inputs |
2 x 8 analogue
inputs
10 V (also as trigger) |
|
Dimensions (LxWxH) |
60' x 60' x 65'
40 x 40 x 78 cm |
|
Weight in kg |
112 (Top) 85
(Bottom) |
|
Physical arrangement |
Two part system |
|
External sampling
system |
Included (with
dilution) |
|
Power requirements |
2 KW, 240 VAC
/ 50 Hz |
|
Optional |
110 V/60 Hz |
|